华西口腔医学杂志 ›› 2016, Vol. 34 ›› Issue (3): 277-280.doi: 10.7518/hxkq.2016.03.012

• 临床研究 • 上一篇    下一篇

3.0 T磁共振6个序列产生金属伪影程度的对比

高岚,廉云敏,王璞,怀海丽   

  1. 河北医科大学第二医院口腔修复科,石家庄050000
  • 收稿日期:2015-10-02 修回日期:2016-03-05 出版日期:2016-06-01 发布日期:2016-06-01
  • 通讯作者: 廉云敏,主任医师,学士,E-mail:lianyunmin@sina.com
  • 作者简介:高岚,副主任医师,学士,E-mail:kuailedegaolan@163.com
  • 基金资助:
    河北省卫生厅科研攻关计划(08115)

Contrastive analysis of artifacts produced by metal dental crowns in 3.0 T magnetic resonance imaging with six sequences

Gao Lan, Lian Yunmin, Wang Pu, Huai Haili   

  1. Dept. of Prosthodontics, The Second Hospital of Hebei Medical University, Shijiazhuang 050000, China) Supported by: Science and Technology Reaserch Plan of Health and Family planning Commission of Hebei(08115).Correspondence: Lian Yunmin, E-mail: lianyunmin@sina.com.
  • Received:2015-10-02 Revised:2016-03-05 Online:2016-06-01 Published:2016-06-01

摘要: 目的 对比3.0 T磁共振检查常用的6种序列所产生金属伪影的程度。方法 制作4种实验铸造冠(钴铬合金、镍铬合金、低钛合金、纯钛),给实验犬依次佩戴4种实验铸造冠,进行3.0 T磁共振自旋回波T1加权像(T1W/SE)、反转恢复T2加权像(T2W/IR)、T2*梯度回波(T2*/GRE)、快速自旋回波T2加权像(T2W/FSE)、液体衰减反转恢复T1加权像(T1W/FLAIR)、螺旋桨技术T2加权像(T2W/PROP)6种序列的头部扫描,测量6种扫描序列扫描4种铸造冠时所产生的伪影最大面积和伪影涉及图像的层数,对测量结果进行对比。结果 从扫描所得的图像及测量数据来看,无论扫描哪种金属,T2*/GRE序列所产生的伪影最大面积和涉及的层数均明显大于其他序列(P<0.01),其他5种序列间差异均无统计学意义(P>0.05)。结论 T2*/GRE序列产生金属伪影程度最大,其余5种序列所产生的金属伪影程度相当。

关键词: 伪影, MRI序列, 快速自旋回波, 自旋回波

Abstract: Objective This study aimed to observe and evaluate six 3.0 T sequences of metallic artifacts produced by metal dental crowns. Methods Dental crowns fabricated with four different materials (Co-Gr, Ni-Gr, Ti alloy and pure Ti) were evaluated. A mature crossbreed dog was used as the experimental animal, and crowns were fabricated for its upper right second premolar. Each crown was examined through head MRI (3.0 T) with six sequences, namely, T1weighted-imaging of spin echo (T1W/SE), T2 weighted-imaging of inversion recovery (T2W/IR), T2 star gradient echo (T2*/GRE), T2 weighted-imaging of fast spin echo (T2W/FSE), T1weighted-imaging of fluid attenuate inversion recovery (T1W/FLAIR), and T2weighted-imaging of propeller (T2W/PROP). The largest area and layers of artifacts were assessed and compared. Results The artifact in the T2*/GRE sequence was significantly wider than those in the other sequences (P<0.01), whose artifact extent was not significantly different (P>0.05). Conclusion T2*/GRE exhibit the strongest influence on the artifact, whereas the five other sequences contribute equally to artifact generation.

Key words: artifacts, MRI sequences, fast spin echo, spin echo

中图分类号: