West China Journal of Stomatology ›› 2016, Vol. 34 ›› Issue (3): 277-280.doi: 10.7518/hxkq.2016.03.012

Previous Articles     Next Articles

Contrastive analysis of artifacts produced by metal dental crowns in 3.0 T magnetic resonance imaging with six sequences

Gao Lan, Lian Yunmin, Wang Pu, Huai Haili   

  1. Dept. of Prosthodontics, The Second Hospital of Hebei Medical University, Shijiazhuang 050000, China) Supported by: Science and Technology Reaserch Plan of Health and Family planning Commission of Hebei(08115).Correspondence: Lian Yunmin, E-mail: lianyunmin@sina.com.
  • Received:2015-10-02 Revised:2016-03-05 Online:2016-06-01 Published:2016-06-01

Abstract: Objective This study aimed to observe and evaluate six 3.0 T sequences of metallic artifacts produced by metal dental crowns. Methods Dental crowns fabricated with four different materials (Co-Gr, Ni-Gr, Ti alloy and pure Ti) were evaluated. A mature crossbreed dog was used as the experimental animal, and crowns were fabricated for its upper right second premolar. Each crown was examined through head MRI (3.0 T) with six sequences, namely, T1weighted-imaging of spin echo (T1W/SE), T2 weighted-imaging of inversion recovery (T2W/IR), T2 star gradient echo (T2*/GRE), T2 weighted-imaging of fast spin echo (T2W/FSE), T1weighted-imaging of fluid attenuate inversion recovery (T1W/FLAIR), and T2weighted-imaging of propeller (T2W/PROP). The largest area and layers of artifacts were assessed and compared. Results The artifact in the T2*/GRE sequence was significantly wider than those in the other sequences (P<0.01), whose artifact extent was not significantly different (P>0.05). Conclusion T2*/GRE exhibit the strongest influence on the artifact, whereas the five other sequences contribute equally to artifact generation.

Key words: artifacts, MRI sequences, fast spin echo, spin echo

CLC Number: