Selection of optimal length and diameter of mini implant in two different forces: a three-dimensional finite element analysis

 Lu Yingjuan, Chang Shaohai, Wu Hong, Yu Yansong, Ye Yushan, Chang Lanru, Wang Wei   

  1. Dept. of Stomatology, Sun Yat-sen Memorial Hospital, Sun Yat-sen University, Guangzhou 510120, China
  • Online:2014-02-01 Published:2014-02-01

Abstract:

 Objective  To investigate the effect of different length and diameters on the stability of mini implant and to select optimal length and diameter using continuous variation of parameters. Methods  To perform 3-dimensional finite element analysis, finite element models of a maxilla, and mini implants with length of 6-12 mm and diameters of 1.2-2.0 mm were generated. Load of two different forces were applied to the head of mini implant. One type was horizontal force (HF), the other was composite force (CF). The maximum equivalent stress (Max EQV) in maxilla and the maximum displacement (Max DM) of mini implant were evaluated. Results  The Max EQV in maxilla and Max DM of mini implant decreased as length and diameter increased. When length was more than 9 mm, the evaluation indexes were small and had a less change. Datas indicated that diameter played a more important role in reducing target, and was a more effective parameter in re-ducing Max EQV when CF was loaded. Conclusion  From biomechanical point of view, the choice of the length should not be more than 9 mm. When CF is loaded using the mini implant, diameter exceeding 1.2 mm are optimal design for mini implant.

Key words: finite element analysis, mini implant, length, diameter, optimization