Thr ee- dimensional finite element analyses of bone sur face str ess of two kinds of conjunction implant

LAN Jing1, XU Xin1, JIANG Guang- shui1, GUAN Yan- jin2, HUANG Hai- yun1, LAN Jing1   

  1. 1. Center of Implantology, School of Stomatology, Shandong University, Jinan 250012, China; 2. Center of Mould and Die Engineering Technology Research, School of Materials Science and Engineering, Shandong University, Jinan 250012, China
  • Received:2008-08-25 Revised:2008-08-25 Online:2008-08-20 Published:2008-08-20
  • Contact: XU Xin,Tel:0531- 88382636

Abstract:

Objective To establish a three- dimension finite element model of mandible with two kinds of dental implant and to study the stress of implant- bone interface. Methods Measuring the data of the components of the dental implant and using spiral CT image reconstruction technique to scan the cross section of the mandible. Threedimension finite element analysis software Unigraphics and MSC. Marc/Mentat were used to build the conjunction model and bone model of two implant systems. Loading 200 N axially and 100 N 30°obliquely on the models respectively, the stress distribution patterns of the bone interface of two implant systems were analyzed. Results The stress distribution on the bone interface of two implant systems was similar. The peak stress of oblique loading was higher than that of axial loading. The peak stress district of the bone was concentrated on the stricture of the implant cervix, which was more obviously displayed on the Replace Select implant. The peak stresses on the bone interface of Replace Select implant were higher than that of Replace implant in all loadings. Conclusion To Replace Select especially, oblique force should be avoided in clinical practice in case of the bone absorption.

Key words: dental implant, finite element analysis, interface